2 ports, single source, base hardware configuration
Cord Power
Qty: 1Wrench Torque 8 in/lb 20 mm Open End
Qty: 1Handle Kit 6-Hole
Qty: 1Keyboard USB
Qty: 1Mouse Optical
Qty: 1Cover Front
Qty: 1Cover Rear
Qty: 1USB Drive with Software and Manuals
Qty: 12 Ports, Single Source, Base Hardware Configuration
Time Domain
Frequency Offset Measurements
Cord Power
Qty: 1Wrench Torque 8 in/lb 20 mm Open End
Qty: 1Cable RF Jumper
Qty: 6Jumper Cable Guard
Qty: 2Handle Kit 6-Hole
Qty: 1Keyboard USB
Qty: 1Mouse Optical
Qty: 1Handle Kit 6-Hole
Qty: 1USB Drive with Software and Manuals
Qty: 1Pulsed RF Measurements
Time Domain
Add pulse modulator to internal first source
Add four internal pulse generators
Frequency-offset measurements
2 ports, single source, base hardware configuration
Cord Power
Qty: 1Wrench Torque 8 in/lb 20 mm Open End
Qty: 1Handle Kit 6-Hole
Qty: 1Keyboard USB
Qty: 1Mouse Optical
Qty: 1Cover Front
Qty: 1Cover Rear
Qty: 1USB Drive with Software and Manuals
Qty: 1Keysight Technologies N52xxA Configuration Guide
Content is AI-generated. Verify critical details with manufacturer-provided resources.
The N5222A with Option 200 operates from 10 MHz to 26.5 GHz with a frequency resolution of 1 Hz. The instrument provides frequency accuracy of ±1 ppm with frequency stability of ±0.05 ppm from -10 to 70 degrees Celsius and ±0.1 ppm per year maximum aging.
The N5222A-200 achieves system dynamic range from 94 dB at 10 to 50 MHz up to 127 dB from 2 to 20 GHz, with typical performance reaching 131 dB in the 2 to 13.5 GHz range. This is calculated as the maximum leveled output power minus the test port noise floor at 10 Hz IF bandwidth.
The N5222A with Option 200 is a 2-port base model with a standard test set. It features 3.5 mm male connectors on test ports and provides simultaneous forward and reverse S-parameter measurements without requiring manual cable reconnections between ports.
The N5222A supports multiple calibration options including the 85052B mechanical calibration kit, N4691B 2-port electronic calibration module, and N4433A 4-port electronic calibration module. Each calibration kit provides different directivity, source match, and load match specifications optimized for various frequency ranges and measurement requirements.
Maximum leveled power varies by frequency: 12 dBm from 10 to 50 MHz, 13 dBm from 50 MHz to 20 GHz, 13 dBm at Port 1 and 10 dBm at Port 2 from 20 to 24 GHz, and 7 dBm at Port 1 and 5 dBm at Port 2 from 24 to 26.5 GHz. Typical performance exceeds these specifications by several dB.
Phase noise performance is frequency-dependent with typical values of -107 dBc/Hz at 1 kHz offset and -127 dBc/Hz at 1 MHz offset for frequencies from 50 MHz to 1 GHz. At higher frequencies from 16 to 26.5 GHz, phase noise is typically -77 dBc/Hz at 1 kHz offset and -92 dBc/Hz at 1 MHz offset.
The N5222A provides adjustable IF bandwidth settings that directly affect measurement speed, noise floor, and dynamic range. Dynamic range specifications are quoted at 10 Hz IFBW, while wider IF bandwidths enable faster measurement throughput at the expense of increased noise floor and reduced dynamic range.
After full 2-port calibration with the 85052B kit at 23 degrees Celsius ±3 degrees, the N5222A achieves directivity of 44 dB, source match of 31 dB, and load match of 44 dB from 2 to 26.5 GHz. Reflection tracking is ±0.006 dB magnitude and ±0.040 degrees phase, while transmission tracking ranges from ±0.091 to ±0.119 dB magnitude across the frequency range.